Pulsed spectrometric amplifier UIS-02
Pulsed spectrometric amplifier UIS-04

To main page / Modular nuclear electronics /

PULSED SPECTROMETRIC AMPLIFIER UIS-04
PURPOSE Increase of image

Amplification and shaping of signals from preamplifiers of semiconductor nuclear-radiation detectors.

USES

The amplifier UIS-04 is used as part of spectrometers (SU-05P, SU-05P1) or with other sources of pulsed signals.

FEATURES
  • Highly accurate pole/zero matching with indication requiring no oscilloscope
  • Active fourth-order filters
  • Gated base line restorer with automatic control
  • Piled-up pulse detector
  • Wide gain range, high linearity, low noise
  • Preamplifier inhibit input with pulsed feedback
  • Generation of INH, BUSY, CRM signals
  • Standard Euromechanics module of width 8 HP (41 mm) and height 3 U (129 mm).

CHARACTERISTICS

Input pulse parameters:
- input pulse polarity +/-
- rise time No more than 200 ns
- decay time constant No less than 50 µ s
Input resistance 1 k Ω
Output pulse parameters:
- output pulse polarity +
- maximum amplitude (at RL = 1 kΩ ) 8.2 V
Quasi-Gauss shaping (T is the shaping time constant) With vertex time 2.2T;
With total duration no longer than 6.4T for the level 0.1
Shaping time constants 1 and 3 µ s or 2 and 6 µ s
Output resistance 50 Ω
Integral nonlinearity No more than В±0.04%
Gain is adjusted by

- “ COARSE” selector switch In steps of 1.277 from 1 to 16
- “ CONTINUOUS” knob From 1 to 1.358
- jumper on the circuit card with a divider 1:1 or 1:10
Basic gain setting error No more than В±20%
Noise level reduced to the input at the gain over 150 No more than 7 µ V at 2 µ s;
No more than 8 µ V at 1 µ s
Intensity characteristics at the input intensity 105 s-1 and shaping time constant 1 µ s or 2 µ s:
- displacement of the center of gravity of the 60Co peak (at least 90% of the full scale) No more than В±1%
- peak broadening No more than 50%