PURPOSE
Registration and transformation of gamma-radiation energy into electric signals
of corresponding amplitude so that they can be later processed.
BDS-G
|
BDS-G100х100
|
BDS-G150х100
|
APPLICATION
- Construction of spectrometric systems for laboratory and industrial purposes.
- Detectors of the BDS-G type can be used in various scientific and economic fields for qualitative and quantitative analysis of gamma-emitting radionuclide samples
COMPONENTS
Detection units are produced on the base of the assembly "scintillation
crystal NaI(TI) - PMT" and consist of:
- pre-amplifier, driver-amplifier with parameters of output impulses that
enable to transmit them directly to the input of the amplitude-digital converter
(multi-channel analyzer)
- high-voltage power supply
- a system for stabilization of the measurement channel based on a special
LED with the function of temperature compensation of converter characteristic
TECHNICAL DATA AND CHARACTERISTICS
Detector type |
BDS-G |
BDS-G 100x100 |
BDS-G 150x100 |
Scintillator size, mm |
63x63 |
100x100 |
150x100 |
Relative energy resolution for the 662-keV line (Cs-137), % |
8 |
9,5 |
10,5 |
Detectable energies, MeV* |
0,03 – 3 |
Integral nonlinearity,% |
1 |
Temperature instability, %/° С |
0,1 |
Additional instability of conversion characteristic at the change of
the load from background to 50000 s-1, % |
1
|
Time instability over a 24-hour period of continuous operation, % |
1
|
Maximum undistorted amplitude of output spectrometric signals, V |
4,8
|
Operating temperatures, °С** |
от +10 до +35 |
Power supply voltage |
6 – 15 |
Power consumption, W |
1,5 |
Weight, kg |
1,4 |
6,5 |
9,3 |
Overall dimensions, mm*** |
Ø 88х315 |
Ø 125х375 |
Ø 180х387 |
NOTES:
* On request, the range of registered energies can be changed
** On request, the range of operating temperatures can be extended from
-20°C to +50°C
*** The size of BDS-G 63 x 63detection part allows the use of Marinelli vessels with capacity
of one litre
|