AMPLITUDE-TO-DIGITAL CONVERTER BPA-03-8K-W1(-SA1) | |
INTERFACE OF BUS compactPCI
Units of format 3U (overall dimensions 160x100 mm) incorporate single-channel spectrometric ADCs and are used in industrial spectrometric systems.
They are made in two versions:
- BPA-03-8K-W1
(foto), Wilkinson ADC, resolution 2K - 8K, software-specified channel width adjusted in steps of 5 µV, input signal 50 mV - 10 V;
- BPA-03-8K-SA1 (foto), ADC with a constant conversion time of 3.5 µs, software-specified
resolution 1K-2K-4K-8K and channel width of 5 mV, 2.5 mV, 1.25 mV, 0.625 mV, input signal 20 mV - 5 V.
GENERAL DATA
Analogue input signal (BNC connector)
|
Polarity
|
Positive
|
Rise time |
< 0.25 µs |
Decay time
|
< 100 µs
|
Input resistance
|
2 k Ω
|
Channel memory capacity
|
224 - 1
|
Real/live time setting accuracy
|
5 ms
|
Ratemeter input - logical signal with frequency
|
< 300 kHz (DB9 connector)
|
Inhibit input with programmable polarity
|
(LEMO connector)
|
BASIC DATA OF BPA-03-8K-SA1
Conversion method
|
Successive approximations with leveling
|
Conversion time
|
3.5 µs
|
Analogue input signal amplitude
|
20 mV - 5 V
|
Software-based selection of resolution (number of channels)
|
1K-2K-4K-8K
|
Differential nonlinearity
|
< 1 %
|
Integral nonlinearity |
< 0.04%
|
Lower-level discriminator threshold range
|
20 mV - 5 V
|
Upper-level discriminator threshold range
|
100 mV - 5 V
|
Spectrum correction by signals from the external pile-up detector
|
+
|
Live time error when piled-up pulse rejection is used:
|
At intensity 100000 pulses/s
|
< 7%
|
At intensity 50000 pulses/s
|
< 4%
|
BASIC DATA OF BPA-03-8K-W1
Conversion method
|
Wilkinson s, 100 MHz
|
Analogue input signal amplitude
|
50 mV - 10 V
|
Resolution (numbers of channels)
|
2K - 8K
|
Channel width, software-adjustable in steps of 5 ОјV
|
2.5 mV - 10 mV |
Differential nonlinearity
|
< 1%
|
Integral nonlinearity
|
< 0.04%
|
Channel width setting accuracy
|
5%
|
Temperature instability of channel width
|
< 0.01%
|
Lower-level discriminator threshold range
|
30 mV - 10 V
|
Upper-level discriminator threshold range |
100 mV - 10 V
|
Print version
|