 | |  | |  | |  | |  | |  | BPA-04 Amplitude to digital converter |  | | |
|
 |
|
/ /
| BPA-04 AMPLITUDE TO DIGITAL CONVERTER | |
FEATURES
- Module in Euromechanics form factor of 3U height and 6HP width;
- Wilkinsons conversion spectrometric ADC with 100 MHz clock frequency;
- Set width of the spectrum channel;
- Internal memory of spectrum accumulation;
- Built-in RS-232, RS-485 and USB interface;
- Multi-channel counting mode
BASIC CHARACTERISTICS
| Conversion method |
Wilkinson |
| Conversion clock frequency |
100 MHz |
| Analogue input signal |
Polarity |
Positive |
| Amplitude |
50 mV 10 V |
| Rise |
> 0.25 μsec |
| Decay |
< 100 μsec |
| Input resistance |
2 kOhm |
| Differential non-linearity (DNL), %, not exceeding |
1 |
| Integral non-linearity (INL), %, not exceeding |
0.04 % |
| Channel width adjustment range (0.005 step), mV |
1.25 to 10 |
| Number of conversion channels |
8 to 1 |
| Channel width setting accuracy, %, not exceeding |
5 % |
| Temperature instability of channel width, %, not exceeding |
0.01 % |
| Channel memory capacity |
224-1 |
| Real/live time setting accuracy |
5 msec |
| Lower-level discriminator threshold range, V |
0.03 to 10 |
| Upper-level discriminator threshold range, V |
0.1 to 10 |
| "Ratemeter" input - logic signal with frequency up to 300 kHz |
+ |
| "Inhibit" input with programmable polarity |
+ |
| Overall dimensions, mm |
19513030 |
| Max weight, g |
500 |
|  |
Print version
|
|