Aspect SPC
About the companyServicesContactsCertificates News     russian  
-Products-
Radiation Monitors and Systems
Dosimetry and radiometry
Nuclear radiation detectors
PC-based MCA
Modular nuclear electronics
Nuclear radiation spectrometers
Inspection systems
Shielding screens
Dedicated equipment
Software

PCI-Wilkinson
PCI-SA
CompactPCI
USB-Wilkinson
USB-SA
BPA-04 Amplitude to digital converter
BPA-04N Amplitude to digital converter

To main page / Modular nuclear electronics / Amplitude-to-digital converters ADC

BPA-04N AMPLITUDE TO DIGITAL CONVERTER

FEATURES
  • Module in NIM form factor of 3U height and 6HP width;
  • Wilkinsons conversion spectrometric ADC with 100 MHz clock frequency;
  • Set width of the spectrum channel;
  • Internal memory of spectrum accumulation;
  • Built-in RS-232, RS-485 and USB interface;
  • Multi-channel counting mode

BASIC CHARACTERISTICS

Conversion method Wilkinson
Conversion clock frequency 100 MHz
Analogue input signal Polarity Positive
Amplitude 50 mV 10 V
Rise > 0.25 μsec
Decay < 100 μsec
Input resistance 2 kOhm
Differential non-linearity (DNL), %, not exceeding 1
Integral non-linearity (INL), %, not exceeding 0.04 %
Channel width adjustment range (0.005 step), mV 1.25 to 10
Number of conversion channels 8 to 1
Channel width setting accuracy, %, not exceeding 5 %
Temperature instability of channel width, %, not exceeding 0.01 %
Channel memory capacity 224-1
Real/live time setting accuracy 5 msec
Lower-level discriminator threshold range, V 0.03 to 10
Upper-level discriminator threshold range, V 0.1 to 10
"Ratemeter" input - logic signal with frequency up to 300 kHz +
"Inhibit" input with programmable polarity +
Overall dimensions, mm 19513030
Max weight, g 500

Print version