Aspect SPC
About the companyServicesContactsCertificates News     russian  
-Products-
Radiation Monitors and Systems
Dosimetry and radiometry
Nuclear radiation detectors
PC-based MCA
Modular nuclear electronics
Nuclear radiation spectrometers
Inspection systems
Shielding screens
Dedicated equipment
Software

scintillation gamma spectrometer Gamma-1S
portable scintillation gamma spectrometer Gamma-1S/NB1
semiconductor gamma spectrometer Gamma-1P
scintillation beta spectrometers Beta-1S, Beta-1S-150
scintillation gamma-beta spectrometer Gamma-Beta-1S
semiconductor alpha spectrometer SEA-13P
radiometer-spectrometer for water radiation control RSKV-01
hand-held radiation monitor MKC-A03
accessories

To main page / Nuclear radiation spectrometers

SEMICONDUCTOR ALPHA SPECTROMETER SEA-13P

PURPOSE
The semiconductor alpha radiation spectrometer SEA-13P is designed for measuring alpha energy spectra, identifying alpha-active nuclides, and determining their concentration in environmental samples prepared by radiochemical separation and directly in samples of aerosol discharges into the atmosphere and production area air samples taken from AFA-RSP-20 type aerosol filters without their radiochemical treatment

SEA-13P. Increase of image

APPLICATION

  • At nuclear-physics centres for research in various fields of fundamental and applied physics
  • At radiochemical laboratories of factories for controlling production processes
  • At laboratories of external radiation monitoring departments of factories for controlling environmental objects
  • at radiological laboratories of the State Sanitary and Epidemiological Inspection, veterinary and agricultural services for controlling he quality of centralized and noncentralized water supply, livestock farming products, and soils

COMPONENTS

  • Measurement chamber unit of the alpha spectrometer with a semiconductor detector (the detector area from 450 to 3000 mm2 is to be ordered by the customer) and with a buil-in ADC (for SEA-13P1)
  • Amplitude-to-digital converter ADC-8К-W1 (for SEA-13P
  • Applied software for Windows -"LSRM"
  • Personal computer with printer
  • Vacuum pump

OPERATION
Operation of the spectrometer is based on conversion of the energy of alpha-particles, which hit the semiconductor detector located in the vacuum chamber, into electric pulses of proportional amplitude with their subsequent recording and analysis

TECHNICAL DATA

Detectable energies, MeV
3,5 – 9,0
Number of analyzer channels
4096
Time to get ready for operation, min
45
Integral nonlinearity in the working range of alpha energies, keV
10
Background in the energy range from 3.5 to 6.0 MeV, pulse/s
0,003
Time instability over the period of continuous operation, keV
10
Maximum statistical input intensity, pulse/s
10000
Continuous operation time, h
8

Characteristic
Rated value
Energy resolution for the 5156-keV line from the 239Pu source of the
OSAI set placed at a distance of 50 mm from the alpha detector of size:
450 mm2
1000 mm2
2000 mm2
3000 mm2


35 keV
50 keV
70 keV
80 keV
Absolute detection efficiency for the 5156-keV line from the 239Pu source
of the OSAI set placed at a distance of 5 mm from the alpha detector of size:
450 mm2
1000 mm2
2000 mm2
3000 mm2


20%
25%
30%
35%

  • The maximum diameter of the sample measured is 70 mm
  • The range of sample-detector distances is 5–50 mm with a step of 5 mm
  • Power consumption, ac 220 V, 50 Hz, is 500 W

OVERALL DIMENSIONS AND WEIGHT OF THE MAIN COMPONENTS

 
Measurement chamber unit
of alpha spectrometer
ADC-8К-W1
(for SEA-13P1)
Vacuum pump NVR-1,25D
Overall dimensions, mm
350х180х250
180х110
136х340х90
Weight, kg
7,5
0,3
10

SOFTWARE
Software ADA (Alpha Decay Analysis)

CERTIFICATES
The instrument is enrolled in the State Register of Measuring Devices under № 15293-96.

Print version

site map e-mail developed by cre8 bureau