PURPOSE
Continuous control of gamma-emitting radionuclides in the running water
APPLICATION
- Water monitoring at waterworks
- Waster water monitoring at NPPs and other plants which use gamma-emitting radionuclides in their technologies
FEATURES
- Continious automatic measurement mode
- High time and temperature stability
- High detection sensitivity
- Alarm if the preset value of the total gamma-activity is exceeded
- Integration into an automated radiation control system
- Creation of data base of spectra and measurement results
- Processing of scintillation gamma spectra
- Identification of radionuclides and calculation of their specific activity
- Certified measurement technique
COMPONENTS
- Measurement chamber with a protection screen
- Scintillation gamma radiation detecting device
- UDS-G3 on the basis of NaI(Ti) 63x160 mm*
- PVC-03 control panel
- Water treatment unit
Note: If the RSKV-01 is used as a part of a radiation control system, the supply set includes necessary communication equipment,
operator's workstation with software for radiation monitoring and spectrum processing software.
TECHNICAL DATA
Detectable energies, MeV |
0,05 3 МэВ |
Relative energy resolution of radiometer against gamma radiation line with energy of 661 keV(Cs-137), % |
9 |
Integral nonlinearity, % |
1 |
Maximal statistic input loading, s-1 |
5 x 104 |
Minimal detected activity of Cs-137 nuclide in a sample during 1 hour measurement time, Bq |
0,5 |
Time instability of conversion characteristic during 24 hours, % |
1 |
Number of radiometer ADC channels |
1024 |
Time necessary for operation mode setting, min |
60 |
Continuous operation time, h |
24 |
Power consumption 220 V, rated supply voltage, VA |
50 |
Operating temperatures, ° С |
5 ÷ 35 |
Overall dimensions:
|
PVC-03 |
400 x 586 x 226 |
measurement chamber |
985 x 585 x 852 |
Weight (radiometer as on assembley), kg |
600 |
NOTES
*Upon request, dimensions of the crystal and the measurement chamber can be modified
OPERATION
The scintillation detecting device detects gamma radiation in the monitored fluid (water) and transfers the results in the form of spectra
to the PVC-03 control panel. Based on these data, the PVC-03 control panel calculates total gamma activity of the monitored water. If
the total gamma activity exceeds the preset threshold, sound and visual alarm is triggered. Measurements are performed in the continuous
automated mode. If the radiometer operates as part of a radiation monitoting system, spectra and processing results are transferred to
the upper-level computer for their detailed processing, including identification and calculation of activity of radionuclides, recording in the
data base and report generation.
The radiometer-spectrometer for water radiation control RSKV-01 (measurement chambler and local control panel PVC-03) is installed directly
at drawoffs. The operator's workstation is installed at the central control point.
CERTIFICATES
The radiometer-spectrometer for water radiation control RSKV-01 is enrolled in the State Register of Measuring Devices under № 22385-02
|