PURPOSE
Qualitative and quantitave analysis of samples for gamma-emittinf radionyclides
APPLICATION
- At dosimetric laboratories, environmental services, radiological laboratories of different ministries and departments to analyze various environmental
samples for presence of gamma-emitting radionuclides
- At radiochemical laboratories to control technological processes
- At nuclear-physics centers to carry out research in various fields of fundamental and applied physics
OPERATION
Operation of the spectrometer is based on conversion of the gamma quantum energy
in the sensitive volume of the semiconductor detector into electric pulses of
a proportional height with their subsequent recording and analysis by the multichannel
pulse-height analyzer.
The spectrometer allows selective and highly sensitive measurement of gamma
radiation spectra in a wide energy range with a high resolution
The spectrometer software allows:
- configuring of the system for various types of detectors
- control of the analyzer parameters, data collecting
- visualization of spectra and their processing results on the computer display
- energy, peak-shape, and efficiency calibration
- automatic processing of gamma spectra with identification of isotopes (by
libraries) and determination of their activity
- storage of spectra and their processing results in files on any information
carrier
TECHNICAL DATA AND CHARACTERISTICS
Detectable energies, MeV |
0,05 - 10 |
Number of channels |
8192 |
Integral nonlinearity, % |
0,05 |
Maximal statistical input intensity, pulse/s |
5 104 |
Energy resolution of the spectrometer (depending on the resolution of
the detector used) for the 60Со 1332-keV gamma line, keV |
1,8 - 3,5 |
Relative detection efficiency at the total gamma absorption peak for the
1332-keV line for a point 60Co nuclide at the source-detector distance 25
cm (depending on the volume of the detector used), % |
10 |
Lowest 137Cs activity measured in a sample for 1 hour with
using a protection screen, Bq |
2 |
Activity measurement error for point geometry when a protection screen
is used, % |
20 |
Power consumption (220V, 50Hz), W |
300 |
Operating temperature, oС |
10...35 |
Protection screen wall thickness: steel 4 mm, lead 82 mm,
copper 1,5 mm, cadmium 1 mm |
OVERALL DIMENSIONS AND WEIGHT OF MAIN SPECTROMETER COMPONENTS
Complete component |
Overall dimensions, mm |
Weight, kg |
Shielding screen and detector with a Dewar vessel |
650х570х1100 |
600 |
Pulse analyzer (computer, ADC, printer) |
Depending on the computer model
|
Spectrometric device SU-03P (single unit) |
375х350х85 |
4,6 |
Spectrometric device SU-04P (NIM crate with modules) |
320х315х235 |
8,9 |
Spectrometric device SU-05P (Euromechanics crate with modules) |
260х250х130 |
4,5 |
FEATURES
The spectrometer hardware and software are easily configured and adapted for
the work with various types of detectors (germanium ones of various energy ranges,
resolutions, and efficiencies; silicon-lithium and silicon ones for detection
of low-energy gamma radiation etc.).
COMPONENTS
- Semiconductor germanium detector (HPGe or Ge(Li))
- Charge-sensitive spectrometric preamplifier PUG-01 (external, for the Ge(Li)
detector)
- Spectrometric device including a spectrometric amplifier, a high-voltage
power supply, and a low-voltage power supply; one of the options below:
Spectrometric
device, standard |
Components |
Detector output voltage ratings |
Preamplifier output voltage ratings |
Input signal parameters |
Output signal parameters |
SU-03P, single unit |
|
Polarity (+/–)
50 4000 V, 100 mА
Automatic step-up/ step-down no more than 15 V/s
|
± 24 V
± 12 V |
Polarity (+/–)
Rise time <200 ns Decay time constant >50 ms |
Polarity (+)
Maximum amplitude 10 V
Forming time constants 0,5; 1; 2;
4; 6; 10 ms |
SU-04P, NIM |
BNN-01 UIS-02 BNV-05 (or BNV-08) |
Polarity (+/–) 50 4000 V, 100 mА; (or 25 1000, 100 mA) Automatic step-up/
step-down no more than 15 V/s
|
± 24 V
± 12 V |
Polarity (+/–)
Rise time <200 ns Decay time constant >50 ms |
Polarity (+)
Maximum amplitude 10 V
Forming time constants 0,5; 1; 2;
4; 6; 10 ms |
SU-05P, Euromechanics |
BNN-03 UIS-04 BNV-07 (or BNV-09)
|
Polarity (+/–) 50 4000 V, 100 mА; (or 5 100 V, 100 mA) or 50 1000 V, 100
mA) Automatic step-up/ step-down no more than 15 V/s
|
± 24 V
± 12 V |
Polarity (+/–)
Rise time <200 ns Decay time constant >50 ms |
Polarity (+)
Maximum amplitude 8,2 V
Forming time constants 2,6 ms |
- Spectrometric amplitude-to-digital converter (ADC) selected for the type
of interface and required characteristics from among the following:
Type of ADC |
Design |
Interface |
Conversion method |
Conversion frequency, MHz (time, μs) |
Number of inputs |
Number of converters |
ADC-8К-2М |
PC circuit card |
|
Wilkinson's |
100 |
2 |
1 |
ADC-8К-W1 |
PC circuit card |
|
|
100 |
1 |
1 |
ADC-8К-W2 |
2 |
2 |
ADC-8К-P1 |
PC circuit card |
|
|
(3,5) |
1 |
1 |
ADC-8К-P2 |
2 |
2 |
BPA-04 |
Euromechanics |
|
Wilkinson's
|
100
|
1 |
1 |
BPA-04N |
NIM |
|
ADC-USB-8К-W |
Stand-alone |
|
|
100 |
1 |
1 |
ADC-USB-8К-P |
Stand-alone |
|
|
(3,5) |
1 |
1 |
ADC-RS-8К-W |
Stand-alone |
RS232/485 |
Wilkinson's |
100 |
1 |
1 |
ADC-RS-8К-P |
Successive
approximations |
(3,5) |
- Personal computer of the IBM PC type with a printer
- Low-background combined lead shielding screen
- Applied software for Windows LSRM2000.
Upon request, the spectrometer can be supplied with the UZ-01 reffiling unit
|